In Vitro Comparison of Surgical Implant Placement Accuracy Using Guides Fabricated by Three Different Additive Technologies

Chuchai Anunmana    
Chananchida Ueawitthayasuporn    
Sirichai Kiattavorncharoen and Prakan Thanasrisuebwong    

Resumen

Various three-dimensional (3D) printing technologies are commercially available on the market, but the influence of different technologies on the accuracy of implant-guided surgery is unclear. Thus, three printing technologies: Stereolithographic (SLA), Digital light processing (DLP), and Polyjet were evaluated in this study. An entire 30 polyurethane models replicated the clinical situation. Ten surgical guides were printed by SLA, DLP, and PolyJet. Then, implant-guided surgery was performed, and their accuracy outcomes were measured concerning angular deviation, 3D deviation at the entry point, and apex. On top of that, the total processing time was also compared. For the angular deviation, the mean deviation was not statistically significant among all technologies. For the 3D deviation, PolyJet was statistically more accurate than DLP (p = 0.002) and SLA (p = 0.017) at the entry point. PolyJet was also statistically more accurate than DLP (p = 0.007) in regards to 3D deviation at the apex. Within the limitation of this study, the deviations from the experiment showed that PolyJet had the best outcome regarding the 3D deviations at the entry point and at the apex, meanwhile, the DLP printer had the shortest processing time.