Performance Test of MicroAeth® AE51 at Concentrations Lower than 2 µg/m3 in Indoor Laboratory

Jeonghoon Lee    

Resumen

It is difficult to find information about how the microAeth® AE51 performs, in spite of its versatility for about a decade in various research fields such as black carbon measurements and personal exposure studies. Stimulated by this, we conducted real-time tests for indoor aerosol in order to provide performance characteristics toward proper usage. We calculated the attenuation (ATN) using the reference signal together with the sensing signal to compare it with the ATN recorded in a microAeth® AE51. Performance was evaluated under extremely low concentration through the zero test, using filtered clean air. Ten-day-long continuous measurements were done for both indoor aerosol and filtered particle free air to examine the feasibility of microAeth® AE51 in an indoor use. Generally, MicroAeth® AE51 exhibited excellent performance, though it showed relatively low performance under some conditions. Noise was intensified while it was directly connected to a power adaptor. Another issue includes the occurrence of negative concentrations for extremely clean air. The noise amplification turned out to be related to a power source independent on the internal battery, and it was able to be removed by post-processing. Uncertainty analysis was carried out to better understand the origin of unwanted noise. Technical perspective of a proper usage will be addressed with regard to what will play a role for a long-term monitoring.